Optimized Marker for Template-Guided Intraoral Surgery

Title
Optimized Marker for Template-Guided Intraoral Surgery
Authors
최진혁김영준이태경정제교김용박세형
Keywords
optimized marker; template-guide; intraoral surgery; registration
Issue Date
2013-02
Publisher
Medicine Meets Virtual Reality 20, Vol 184, 2013
Citation
VOL 184, 85-91
Abstract
Template-guided intraoral surgery is attracting interest due to its accuracy and convenience. Usually, fiducial markers are used for registration of the different model coordinates for patients or plaster and fabrication coordinates in template-guided dental surgery. The accuracy of markers is important because it is directly related to the accuracy of the template-guide, which ultimately affects the success of surgical results. In this paper, we present the methods of determining optimal marker parameters such as size, shape, and material. We have defined parameters that could potentially influence the accuracy of markers and tested them with various values for each parameter. Experiments were performed to measure repeatability, and we tested accuracy and minimized potential error by comparing the actual and calculated dimensions. As a result, the following optimal marker parameters were identified: 3-mm diameter, positively tapered cylindrical shape, and titanium composition.
URI
http://pubs.kist.re.kr/handle/201004/44559
ISSN
09269630
Appears in Collections:
KIST Publication > Conference Paper
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE