The effects of the size of a boiling bubble on lesion production in boiling histotripsy

Title
The effects of the size of a boiling bubble on lesion production in boiling histotripsy
Authors
김형민박기주M O de AndradeN Saffari
Keywords
High Intensity Focused Ultrasound; Boiling Histotripsy; Bubble Dynamics; Cavitation
Issue Date
2019-04
Publisher
Journal of Physics: Conference Series
Citation
VOL 1184-9
Abstract
Boiling histotripsy employs a number of millisecond-long High Intensity Focused Ultrasound (HIFU) pulses with high acoustic peak pressures at the HIFU focus to mechanically fractionate soft tissue. Studies have shown the mechanisms underpinning this tissue fractionation process; however, the question of how HIFU exposure conditions affect lesion formation still remains unclear. In the present work, a high-speed camera and a passive cavitation detection (PCD) system were used to investigate the dynamics of bubbles induced and the corresponding mechanical damage generated in optically transparent tissue-mimicking phantoms with two different boiling histotripsy exposure conditions (1. P+ = 85.4 MPa, P- = – 15.6 MPa; 2. P+ = 71.5 MPa, P- = – 13.4 MPa at focus). Our results clearly showed that there is a positive relationship between the size of a boiling bubble and the lesion dimension. At P+ = 85.4 MPa and P- = – 15.6 MPa, a relatively larger boiling bubble was, for instance, produced at the focus in the gel phantom followed by the presence of a wider cavitation cluster progressing toward the HIFU transducer, resulting in the formation of a larger lesion compared to that with P+ = 71.5 MPa and P- = – 13.4 MPa.
URI
http://pubs.kist.re.kr/handle/201004/69205
ISSN
1742-6596
Appears in Collections:
KIST Publication > Article
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML


qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE