Fast Myoglobin Detection Using Nanofluidic Electrokinetic Trapping Technique
- Authors
- Chun, DongWon; Kim, Sang Hui; Song, Hyungwan; Kwak, Seungmin; Kim, YooChan; Seok, HyunGwang; Lee, Sang-Myung; Lee, Jeong Hoon
- Issue Date
- 2013-01
- Publisher
- JAPAN SOC APPLIED PHYSICS
- Citation
- APPLIED PHYSICS EXPRESS, v.6, no.1
- Abstract
- We report on the preconcentration-enhanced fast collection of myoglobin protein for the rapid detection of myocardial infarction. We use a one-dimensional micro/nanofluidic chip for electrokinetic preconcentration and demonstrate that the preconcentration factor of 1 ng/ml Alexa Fluor 488-labeled myoglobin is similar to 1000 within 200 s, where the protein had a weak negative charge, thereby making it hard to perform electrokinetic trapping for neutral-like proteins. The potential feasibility with new assay strategies for use in a rapid immunoassay screening test for myocardial infarction is discussed. (C) 2013 The Japan Society of Applied Physics
- Keywords
- PRECONCENTRATION; BIOMARKERS; MEMBRANE; PROTEINS; PRECONCENTRATION; BIOMARKERS; MEMBRANE; PROTEINS; Myoglobin detection; nanofluidics; Electrokinetic trapping
- ISSN
- 1882-0778
- URI
- https://pubs.kist.re.kr/handle/201004/128481
- DOI
- 10.7567/APEX.6.017001
- Appears in Collections: