인공고관절 전치환술에서 세라믹 볼 헤드의 기계적 안정성 평가를 위한 유한요소 해석
- Other Titles
- Mechanical safety evaluation of ceramic ball head for total hip replacement using finite element method
- Authors
- 한성민; 추준욱; 전흥재; 김정성; 최귀원; 윤인찬
- Issue Date
- 2010-12
- Publisher
- 대한의용생체공학회
- Citation
- 의공학회지, v.31, no.6, pp.449 - 455
- Abstract
- A ceramic articulating system in total hip replacement thought to be superior to metal-on-polyethylene due to its extremely low coefficient of friction and potential for high resistance to wear. But ceramic is brittle, which makes it mechanically and theoretically susceptible to fracture under certain mechanical conditions. In the current study, nine different models of ceramic ball heads were mechanically evaluated using 3D finite element(FE) analyses. It was found that the maximum stress in all ceramic models was lower than ceramic flexural strength, and it satisfied the requirements of the FDA Gaudience for artificial hip implant. Thus, ceramic ball head models introduced in the current study could be mechanically safe for clinical applications.
- Keywords
- Total Hip Replacement(THR); ceramic ball head; Finite Element Method(FEM); mechanical safety
- ISSN
- 1229-0807
- URI
- https://pubs.kist.re.kr/handle/201004/130841
- Appears in Collections:
- KIST Article > 2010
- Files in This Item:
There are no files associated with this item.
- Export
- RIS (EndNote)
- XLS (Excel)
- XML