Label free novel electrical detection using micromachined PZT monolithic thin film cantilever for the detection of C-reactive protein

Authors
Lee, JHYoon, KHHwang, KSPark, JAhn, SKim, TS
Issue Date
2004-09-15
Publisher
ELSEVIER ADVANCED TECHNOLOGY
Citation
BIOSENSORS & BIOELECTRONICS, v.20, no.2, pp.269 - 275
Abstract
In this paper, we report the novel electrical measurement for the label-free detection of C-reactive protein (CRP) using resonant frequency shift in the monolithic thin film cantilever of micromachined Pb(Zr0.52Ti0.48)O-3 (PZT) which was fabricated with the composition of SiO2/Ta/Pt/PZT/Pt/SiO2 on silicon nitride (SiNx) supporting layer for the dual purpose of electrical self-excitation and sensing. The specific binding characteristics of CRP antigen to its antibody, which is immobilized with Calixcrown SAMs on An surface deposited on microcantilever, is determined in high sensitivity to the nanogram level per milliliter by measuring the resonant frequency shift. The nanomechanical PZT cantilever turns out a robust platform for the highly specific antigen-antibody interaction and provides with the novel tool for qualification and quantification of biomolecules without any sample labeling and bulky optical apparatus. (C) 2004 Elsevier B.V. All rights reserved.
Keywords
SURFACE STRESS; ALKANETHIOLS; SURFACE STRESS; ALKANETHIOLS; nanomechanical biosensor; cantilever; PZT film; CRP; resonant frequency shift
ISSN
0956-5663
URI
https://pubs.kist.re.kr/handle/201004/137226
DOI
10.1016/j.bios.2004.01.024
Appears in Collections:
KIST Article > 2004
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE