실시간 미세입자 측정을 위한 다이오드형 코로나 하전기의 설계 및 성능평가

Other Titles
Design and Performance Evaluation of a Diode Type Corona Charger for Real-Time Measurement of the Submicron Aerosol
Authors
황정호배귀남조명훈지준호박동호
Issue Date
2004-09
Publisher
대한기계학회
Citation
대한기계학회논문집 B, v.28, no.9, pp.1066 - 1074
Abstract
With a diode corona charger, which is a component of ELPI(Electrical Low Pressure Impactor), aerosol particles are charged to make electrical detection possible before they are collected by the impactor. We designed and evaluated two cylindrical corona chargers, each of which had a central corona needle electrode. For the performance evaluation of each corona charger the polydisperse dioctyl sebacate(DOS) particles, with diameters of 0.1~0.8 mm, and NaCl particles, smaller than 0.1 mm, were used. The particles were then led through an electrostatic classifier (TSI model 3081) to classify monodisperse aerosol with minimal size deviation. After evaluating the wall loss of the particles in the corona charger, we measured the product of penetration and number of charges, , to evaluate the corona charger efficiency at high positive voltages of 4, 5, 6 kV.
Keywords
Electrical Low Pressure Impactor(전기적 저압 임팩터); Real-Time(실시간); Corona Charger(코로나 하전기); Monodisperse Aerosol(단분산 입자); Electrical Low Pressure Impactor(전기적 저압 임팩터); Real-Time(실시간); Corona Charger(코로나 하전기); Monodisperse Aerosol(단분산 입자)
ISSN
1226-4881
URI
https://pubs.kist.re.kr/handle/201004/137258
Appears in Collections:
KIST Article > 2004
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.