아질산성질소 측정용 FIA의 제작 및 응용에 관한 연구(광주광역시 광주천 시료를 대상으로)

Other Titles
A Study on the Constitution and the Application of FIA System for Measurement of Nitrite (The Field Water Samples at Kwangju)
Authors
이재성박완철이수원김용준
Issue Date
2002-05
Publisher
한국물환경학회
Citation
한국물환경학회지, v.18, no.3, pp.283 - 290
Abstract
In this study, home-made detection system by means of noble FIA is introduced on the measurement of toxic nitrite in the water samples collected from the area of Kwangju. As the standard calibration between 30 to 1000 ppb, the linearity has been shown more than 0.9999 as the correlation coefficient(R2) with the detection limit 1.5 ppb(S/N>2). The distribution of sample concentration was monitored as N.D. - 123 ppb which is wide span of concentrations in field water samples. The low level of nitrite is hardly detectable with other expensive sophisticated instruments including ion chromatography. Whereas the result of high concentration brings forth the necessity monitoring constantly our precious water resources. Successfully, the FIA system has played a very important role detecting wide span of nitrite in water sample. This technique can be adopted for controlling our environment in the near future.
Keywords
FIA; water monitoring; nitrite; automatic detection; FIA; water monitoring; nitrite; automatic detection