Dielectric hysteresis measurement in lossy ferroelectrics

Authors
Park, JHKim, BKPark, JGKim, ITJe, HJKim, YPark, SJ
Issue Date
1999-12
Publisher
GORDON BREACH SCI PUBL LTD
Citation
FERROELECTRICS, v.230, no.1-4, pp.453 - 458
Abstract
The conventional Sawyer-Tower circuit is modified so as to employ a quasi-DC electric-field rather than the 60Hz AC electric-field. By using this modified Sawyer-Tower method, the P-E hysteresis loop can be determined using 1 cycle of electric field, This technique enables the samples to avoid the joule heating induced breakdown, which often occurred under the cyclic load of the 60 Hz AC electric-field. As the quasi-DC electric-field was used in this study, it is very important to compensate the phase shift by the lossy conduction during measurement. A novel method of compensation is also to be presented.
Keywords
polarization; P-E hysteresis loop