An FPGA-Based Embedded System for Portable and Cost-Efficient Bio-sensing A Low-Cost Controller for Biomedical Diagnosis
- Authors
- Lee, Ikho; Kim, Byungsub; Yi, Hyunjung; Lee, Ki-Young; Lee, Seung-Woo
- Issue Date
- 2015-09
- Publisher
- IEEE
- Citation
- IEEE International Circuits and Systems Symposium (ICSyS 2015), pp.1 - 6
- Abstract
- This paper presents a design of an embedded system for a portable and cost effective bio-sensing. With our system, users can easily and cost-efficiently generate the necessary control signals such as cyclic waves for bio-sensors with precisely controlled timing. The bio-sensing system consists of an analog-to-digital converter (ADC) used for data acquisition, digital-to-analog converter (DAC) in order to generate cyclic wave for a biosensor, field programmable gate array (FPGA) to interface with computer, and potentiostat for biosensor signal processing. Our system has two modes that are software controlled real time mode for real-time data acquisition and hardware controlled burst mode for high-speed data acquisition. The system accuracy was evaluated by series R-test. In cyclic voltammetry test with a Ru(NH3)(6)Cl-3, a graph provided by our system has 2.77% relative error at current peak compared with the data acquired by a 18 times more expensive commercial system.
- URI
- https://pubs.kist.re.kr/handle/201004/115015
- Appears in Collections:
- KIST Conference Paper > 2015
- Files in This Item:
There are no files associated with this item.
- Export
- RIS (EndNote)
- XLS (Excel)