Sensitivity of GC-EI/MS, GC-EI/MS/MS, LC-ESI/MS/MS, LC-Ag+CIS/MS/MS, and GC-ESI/MS/MS for analysis of anabolic steroids in doping control
- Authors
- Cha, Eunju; Kim, Sohee; Kim, Ho Jun; Lee, Kang Mi; Kim, Ki Hun; Kwon, Oh-Seung; Lee, Jaeick
- Issue Date
- 2015-11
- Publisher
- WILEY-BLACKWELL
- Citation
- DRUG TESTING AND ANALYSIS, v.7, no.11-12, pp.1040 - 1049
- Abstract
- This study compared the sensitivity of various separation and ionization methods, including gas chromatography with an electron ionization source (GC-EI), liquid chromatography with an electrospray ionization source (LC-ESI), and liquid chromatography with a silver ion coordination ion spray source (LC-Ag+CIS), coupled to a mass spectrometer (MS) for steroid analysis. Chromatographic conditions, mass spectrometric transitions, and ion source parameters were optimized. The majority of steroids in GC-EI/MS/MS and LC-Ag+CIS/MS/MS analysis showed higher sensitivities than those obtained with other analytical methods. The limits of detection (LODs) of 65 steroids by GC-EI/MS/MS, 68 steroids by LC-Ag+CIS/MS/MS, 56 steroids by GC-EI/MS, 54 steroids by LC-ESI/MS/MS, and 27 steroids by GC-ESI/MS/MS were below cut-off value of 2.0 ng/mL. LODs of steroids that formed protonated ions in LC-ESI/MS/MS analysis were all lower than the cut-off value. Several steroids such as unconjugated C3-hydroxyl with C17-hydroxyl structure showed higher sensitivities in GC-EI/MS/MS analysis relative to those obtained using the LC-based methods. The steroids containing 4, 9, 11-triene structures showed relatively poor sensitivities in GC-EI/MS and GC-ESI/MS/MS analysis. The results of this study provide information that may be useful for selecting suitable analytical methods for confirmatory analysis of steroids. Copyright (c) 2015 John Wiley & Sons, Ltd.
- Keywords
- ATMOSPHERIC-PRESSURE PHOTOIONIZATION; TANDEM MASS-SPECTROMETRY; LIQUID-CHROMATOGRAPHY; HUMAN URINE; CHEMICAL-IONIZATION; ANDROGENIC STEROIDS; SCREENING METHOD; CONFIRMATION; AGENTS; anabolic steroids; gas chromatography-mass spectrometry; liquid chromatography-tandem mass spectrometry; sensitivity; doping control; ionization
- ISSN
- 1942-7603
- URI
- https://pubs.kist.re.kr/handle/201004/124806
- DOI
- 10.1002/dta.1906
- Appears in Collections:
- KIST Article > 2015
- Files in This Item:
There are no files associated with this item.
- Export
- RIS (EndNote)
- XLS (Excel)
- XML
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.