BioFET sensor for detection of albumin in urine
- Authors
- Park, K. -M.; Lee, S. -K.; Sohn, Y. -S.; Choi, S. -Y.
- Issue Date
- 2008-01-31
- Publisher
- INST ENGINEERING TECHNOLOGY-IET
- Citation
- ELECTRONICS LETTERS, v.44, no.3, pp.185 - U7
- Abstract
- Characterisation of a BioFET for detection of albumin in a mixture of human urine is presented. To avoid electrolyte effect of the urine, it was measured in PBS (phosphate buffer saline) at a fixed pH after albumin binding. The drain current was modulated by the albumin bound to the anti-albumin immobilised on the gate surface of the BioFET. The current variation ratio was likely to be proportional to the concentration of the albumin in the range 50 - 250 mg/1. The results show the feasibility of the BioFET as a urinary albumin sensor.