Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | 황정호 | - |
dc.contributor.author | 배귀남 | - |
dc.contributor.author | 조명훈 | - |
dc.contributor.author | 박동호 | - |
dc.contributor.author | 지준호 | - |
dc.date.accessioned | 2024-01-21T06:42:29Z | - |
dc.date.available | 2024-01-21T06:42:29Z | - |
dc.date.created | 2021-09-06 | - |
dc.date.issued | 2004-07 | - |
dc.identifier.issn | 1226-4881 | - |
dc.identifier.uri | https://pubs.kist.re.kr/handle/201004/137438 | - |
dc.description.abstract | Cascade impactors are widely used to collect size classified aerosol. A major disadvantage of this instrument is the required long sampling time. Electrical low pressure impactor has been developed to overcome this disadvantage and to achieve real-time measurements on the particle size distribution. The instrument consists primarily of a corona charger, low pressure cascade impactor and multi channel electrometer. We designed and evaluated the performance of a potable 3-stage low pressure impactor using an electrical method. For the calibration of the impactor, monodispersed particles were generated using evaporation-condensation method followed by electrostatic classification using a DMA(Differential Mobility Analyzer). The collection efficiency curves of the stages can be determined by analysing the fraction of particles collected by each stage. | - |
dc.publisher | 대한기계학회 | - |
dc.title | 미세입자 측정용 간이형 3단 전기적 저압 임팩터의 설계 및 성능평가 | - |
dc.title.alternative | Design and Performance Evaluation of a Portable 3-Stage Electrical Low Pressure Impactor(P-ELI) for Measurements of Submicron Aerosol | - |
dc.type | Article | - |
dc.description.journalClass | 2 | - |
dc.identifier.bibliographicCitation | 대한기계학회논문집 B, v.28, no.7, pp.826 - 833 | - |
dc.citation.title | 대한기계학회논문집 B | - |
dc.citation.volume | 28 | - |
dc.citation.number | 7 | - |
dc.citation.startPage | 826 | - |
dc.citation.endPage | 833 | - |
dc.description.journalRegisteredClass | kci | - |
dc.identifier.kciid | ART001095262 | - |
dc.subject.keywordAuthor | Electrical Low Pressure Impactor(전기적 저압 임팩터) | - |
dc.subject.keywordAuthor | Cascade Impactor(다단 임팩 터) | - |
dc.subject.keywordAuthor | Cutsize(분리입경) | - |
dc.subject.keywordAuthor | Stokes Number(스톡스 수) | - |
dc.subject.keywordAuthor | Electrical Low Pressure Impactor(전기적 저압 임팩터) | - |
dc.subject.keywordAuthor | Cascade Impactor(다단 임팩 터) | - |
dc.subject.keywordAuthor | Cutsize(분리입경) | - |
dc.subject.keywordAuthor | Stokes Number(스톡스 수) | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.