Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kang, Min-Jung | - |
dc.contributor.author | Lee, Hyeim | - |
dc.contributor.author | Shin, Sang Mi | - |
dc.contributor.author | Yoo, Hye Hyun | - |
dc.contributor.author | Kwon, Oh-Seung | - |
dc.contributor.author | Jin, Changbae | - |
dc.contributor.author | Chung, Bong Chul | - |
dc.date.accessioned | 2024-01-20T18:32:00Z | - |
dc.date.available | 2024-01-20T18:32:00Z | - |
dc.date.created | 2021-09-05 | - |
dc.date.issued | 2010-10 | - |
dc.identifier.issn | 0253-2964 | - |
dc.identifier.uri | https://pubs.kist.re.kr/handle/201004/131044 | - |
dc.description.abstract | The objective of this study was to validate performance characteristics of the Access 2 (Beckman coulter) system for hCG assays for use as a confirmation test for doping control. The Access 2 assay was linear up to 500 IU/L. The correlation coefficient was higher than 0.999, and the sensitivity of the linearity was 0.82. There were no false positive or false negative cases. LOD was 0.59 IU/L. The method was robust when performed by different people. Repeatability and reproducibility were below 7%. We compared reproducibility and recoveries of Access 2 and Elecsys 2010. Access 2 demonstrated higher reproducibility than Elecsys 2010. Recoveries (accuracy) of Access 2 were between 85 and 105%. Recoveries for Elecsys 2010 were between 91 and 104%. | - |
dc.language | English | - |
dc.publisher | 대한화학회 | - |
dc.title | Performance Characteristic of the beta-Human Chorionic Gonadotropin Assay on the Access 2 for Its Use in Doping Control | - |
dc.type | Article | - |
dc.identifier.doi | 10.5012/bkcs.2010.31.10.2889 | - |
dc.description.journalClass | 1 | - |
dc.identifier.bibliographicCitation | Bulletin of the Korean Chemical Society, v.31, no.10, pp.2889 - 2892 | - |
dc.citation.title | Bulletin of the Korean Chemical Society | - |
dc.citation.volume | 31 | - |
dc.citation.number | 10 | - |
dc.citation.startPage | 2889 | - |
dc.citation.endPage | 2892 | - |
dc.description.isOpenAccess | N | - |
dc.description.journalRegisteredClass | scie | - |
dc.description.journalRegisteredClass | scopus | - |
dc.description.journalRegisteredClass | kci | - |
dc.identifier.kciid | ART001526141 | - |
dc.identifier.wosid | 000288766000022 | - |
dc.identifier.scopusid | 2-s2.0-78049382860 | - |
dc.relation.journalWebOfScienceCategory | Chemistry, Multidisciplinary | - |
dc.relation.journalResearchArea | Chemistry | - |
dc.type.docType | Article | - |
dc.subject.keywordPlus | TESTOSTERONE | - |
dc.subject.keywordPlus | SPORTS | - |
dc.subject.keywordPlus | MISUSE | - |
dc.subject.keywordAuthor | beta-hCG | - |
dc.subject.keywordAuthor | Access 2 | - |
dc.subject.keywordAuthor | Doping control | - |
dc.subject.keywordAuthor | Method evaluation | - |
dc.subject.keywordAuthor | Confirmation method | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.