Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Choo, Sung Joong | - |
dc.contributor.author | Lee, Byung-Chul | - |
dc.contributor.author | Lee, Sang-Myung | - |
dc.contributor.author | Park, Jung Ho | - |
dc.contributor.author | Shin, Hyun-Joon | - |
dc.date.accessioned | 2024-01-20T21:00:45Z | - |
dc.date.available | 2024-01-20T21:00:45Z | - |
dc.date.created | 2021-09-04 | - |
dc.date.issued | 2009-09 | - |
dc.identifier.issn | 0960-1317 | - |
dc.identifier.uri | https://pubs.kist.re.kr/handle/201004/132211 | - |
dc.description.abstract | In this paper, silicon oxynitride layers deposited with different plasma-enhanced chemical vapor deposition (PECVD) conditions were fabricated and optimized, in order to make an interferometric sensor for detecting biochemical reactions. For the optimization of PECVD silicon oxynitride layers, the influence of the N(2)O/SiH(4) gas flow ratio was investigated. RF power in the PEVCD process was also adjusted under the optimized N(2)O/SiH(4) gas flow ratio. The optimized silicon oxynitride layer was deposited with 15 W in chamber under 25/150 sccm of N(2)O/SiH(4) gas flow rates. The clad layer was deposited with 20 W in chamber under 400/150 sccm of N(2)O/SiH(4) gas flow condition. An integrated Mach-Zehnder interferometric biosensor based on optical waveguide technology was fabricated under the optimized PECVD conditions. The adsorption reaction between bovine serum albumin (BSA) and the silicon oxynitride surface was performed and verified with this device. | - |
dc.language | English | - |
dc.publisher | Institute of Physics Publishing | - |
dc.title | Optimization of silicon oxynitrides by plasma-enhanced chemical vapor deposition for an interferometric biosensor | - |
dc.type | Article | - |
dc.identifier.doi | 10.1088/0960-1317/19/9/095007 | - |
dc.description.journalClass | 1 | - |
dc.identifier.bibliographicCitation | Journal of Micromechanics and Microengineering, v.19, no.9 | - |
dc.citation.title | Journal of Micromechanics and Microengineering | - |
dc.citation.volume | 19 | - |
dc.citation.number | 9 | - |
dc.description.isOpenAccess | N | - |
dc.description.journalRegisteredClass | scopus | - |
dc.identifier.wosid | 000269474500027 | - |
dc.identifier.scopusid | 2-s2.0-70350680938 | - |
dc.relation.journalWebOfScienceCategory | Engineering, Electrical & Electronic | - |
dc.relation.journalWebOfScienceCategory | Nanoscience & Nanotechnology | - |
dc.relation.journalWebOfScienceCategory | Instruments & Instrumentation | - |
dc.relation.journalWebOfScienceCategory | Physics, Applied | - |
dc.relation.journalResearchArea | Engineering | - |
dc.relation.journalResearchArea | Science & Technology - Other Topics | - |
dc.relation.journalResearchArea | Instruments & Instrumentation | - |
dc.relation.journalResearchArea | Physics | - |
dc.type.docType | Editorial Material | - |
dc.subject.keywordPlus | OPTICAL SENSORS | - |
dc.subject.keywordPlus | WAVE-GUIDES | - |
dc.subject.keywordPlus | FABRICATION | - |
dc.subject.keywordPlus | TECHNOLOGY | - |
dc.subject.keywordAuthor | Biosensor | - |
dc.subject.keywordAuthor | MEMS | - |
dc.subject.keywordAuthor | Planar waveguide | - |
dc.subject.keywordAuthor | PECVD | - |
dc.subject.keywordAuthor | interferometric sensor | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.