새로운 장치를 이용한 인공혈관 다공성에 관한 연구
- Other Titles
- Evaluation of Vascular Graft Porosity Using Newly Designed Device
- Authors
- 이세철; 이신실; 김규석; 조재화; 윤창진; 이태승; 강성권; 박재형; 김수현
- Issue Date
- 2009-04
- Publisher
- 한국생체재료학회
- Citation
- 생체재료학회지, v.13, no.2, pp.55 - 59
- Abstract
- To establish quantitative analysis of the porosity of graft, this research had developed a permeability measuring
machine. And the machine compared a weaving graft, firstly developed graft in Korea, with other two existing weaving
grafts-made by the Kitamura and the Nakao. The tests for permeability of three different grafts were conducted with
saline and animal blood. And the pressures were 120 mmHg and 200 mmHg. With saline, the newly developed graft
resulted in 327 and 268(ml/cm2/min) for each pressure, and the Kitamura’s were 249 and 305, and the Nakao°Øs
were 244 and 303. And also with animal blood, the new graft marked 202 and 246(ml/ cm2/min) for each pressure,
and the Kitamura marked 180 and 215, also the Nakao marked 193 and 220. As the result, the comparison data with
actual measurement which were calculated mathematically by the porosity and the specific surface area of the three
grafts are indicating same tendency. Therefore, the permeability measuring machine can be said that it is effective in
establishing the quantitative analysis of porosity of a graft.
- Keywords
- Vascular graft; Graft porosity; Graft mechanical properties
- ISSN
- 1226-4601
- URI
- https://pubs.kist.re.kr/handle/201004/132590
- Appears in Collections:
- KIST Article > 2009
- Files in This Item:
There are no files associated with this item.