A Multimodal Multichannel Neural Activity Readout IC with 0.7 mu W/Channel Ca2+-Probe-Based Fluorescence Recording and Electrical Recording
- Authors
- Lee, Taeju; Park, Jee-Ho; Cha, Ji-Hyoung; Chou, Namsun; Je, Minkyu; Kim, Ji-Hoon; Cho, Il-Joo; Kim, Seong-Jin; Jang, Doojin
- Issue Date
- 2019-06
- Publisher
- IEEE
- Citation
- Symposium on VLSI Technology / 33rd Symposium on VLSI Circuits, pp.C290 - C291
- Abstract
- This paper presents a multimodal multichannel neural activity readout IC which can perform not only the electrical recording (ER) but also the fluorescence recording (FR) of neural activity for the cell-type-specific study of heterogeneous neuronal cell populations. The time-based FR circuit senses Ca2+ concentration using Ca(2+)probes while the ER circuit acquires action potentials (APs) and local field potentials (LFPs). The IC is fabricated in 0.18 mu m CMOS, The FR circuit achieves a recording range of 81dB (75pA to 860nA) and consumes the power of 0.7 mu W/Ch. The ER circuit achieves the input-referred noise (IRN) of 2.7 mu V-rms over the bandwidth (BW) of 10kHz, while consuming the power of 4.9 mu W/Ch. The in-vitro measurement is performed for recording Ca2+ concentration and electrical neural signals.
- URI
- https://pubs.kist.re.kr/handle/201004/114084
- Appears in Collections:
- KIST Conference Paper > 2019
- Files in This Item:
There are no files associated with this item.