Calibration method for microscale stereo X-ray imaging system
- Authors
- Kim, Youngjun; Kim, Wontae; Park, Sehyung; Kim, Jong-Hyeong
- Issue Date
- 2012-06
- Publisher
- KOREAN SOC PRECISION ENG
- Citation
- INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, v.13, no.6, pp.877 - 882
- Abstract
- Calibration of an X-ray imaging system involves estimation of the 3D geometrical configuration of the system components. Here, we propose a calibration method for stereo X-ray imaging systems, which capture two X-ray images at two orthogonal positions. The calibration parameters include the relation between image coordinates and world coordinates, the relative 3D positions of the X-ray source and detector, and the rotational axis of the stereo X-ray system. The average error of the proposed calibration method was determined to be approximately 0.03% in evaluation tests.
- Keywords
- ATTENUATION FIELDS; RADIOGRAPHS; GENERATION; ATTENUATION FIELDS; RADIOGRAPHS; GENERATION; Calibration; Dual X-ray; Micro-scale X-ray; Radiographic imaging; Stereo X-ray
- ISSN
- 2234-7593
- URI
- https://pubs.kist.re.kr/handle/201004/129203
- DOI
- 10.1007/s12541-012-0114-3
- Appears in Collections:
- KIST Article > 2012
- Files in This Item:
There are no files associated with this item.
- Export
- RIS (EndNote)
- XLS (Excel)
- XML