Homogenized stress analysis in a dental implant system

Authors
Moon, HJLee, JHChoi, KChoi, JBKoh, CS
Issue Date
1997-11
Publisher
TAYLOR & FRANCIS LTD
Citation
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, v.21, no.6, pp.233 - 240
Abstract
In this study, the stress state occurring at each trabeculae due to three different types of dental implant is investigated by a homogenization technique, in which the trabecular structure is assumed to be composed of repeating hexagonal units. This technique helps make a proper material model of bone and to analyse such a non-homogeneous structure at the level of an individual microstructural unit. Stress analyses with the homogenization technique show a much higher stress level in the sponge bone, compared to those with conventional FEM. It also shows that even a minor lateral force results in crucial stresses in the dental implant system. The stress states of the mandible with a hemisphere-rooted implant and a wedge type implant show similar levels, while those with a rectangular-rooted implant result in higher stresses. It is suggested that the distance between the implant tip and cortical bone be kept far enough apart to prevent stress concentrations in the mandible.
ISSN
0309-1902
URI
https://pubs.kist.re.kr/handle/201004/143528
DOI
10.3109/03091909709070014
Appears in Collections:
KIST Article > Others
Files in This Item:
There are no files associated with this item.
Export
RIS (EndNote)
XLS (Excel)
XML

qrcode

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

BROWSE