The optimization of ELISA for methamphetamine determination: The effect of immunogen, tracer and antibody purification method on the sensitivity

Authors
Choi, JChoi, MJKim, CCho, YSChin, JJo, YA
Issue Date
1997-02
Publisher
PHARMACEUTICAL SOCIETY KOREA
Citation
ARCHIVES OF PHARMACAL RESEARCH, v.20, no.1, pp.46 - 52
Abstract
To obtain more sensitive immunoassay for methamphetamine (MA) determination, the optimum condition of enzyme-linked immunosorbent assay (ELISA) was investigated in regard to immunogens, antibody purification methods and coating tracers. Activated MA, N-(4-aminobutyl)methamphetamine (4-ABMA), was conjugated with bovine serum albumin (BSA) or keyhole limpet hemocyanin (KLH) and used as immunogen. The antibodies were purified by protein G chromatography or various immunoaffinity chromatography-linked MA-protein ligands, such as MA-BSA, MA-KLH or MA-ovalbumin (OVA). Each purified antibody was characterized by means of sensitivity and cross-reactivity using the three MA-protein coating tracers, MA-BSA, MA-KLH and MA-OVA. The best sensitivity of each antibody was acquired with the MA-OVA tracer although the tracer concentration and the antibody titer level at optimum condition were varied. The antibody with high titer level did not always yield good sensitivity. At optimum condition, immunoaffinity chromatography-purified antibodies were better for sensitivity and for specificity than protein C-purified antibodies. The cross-reactivity of the purified antibodies seemed to be affected by immunogen structure and showed somewhat different patterns according to the immunoaffinity ligand utilized. These data show that the antibody purification method as well as choice of coating tracer and immunogen is essential for the sensitivity and specificity of EIA; the optimum condition for assay should be discovered using various methods and combinations.
Keywords
POLARIZATION FLUOROIMMUNOASSAY; AMPHETAMINE; IMMUNOASSAY; URINE; POLARIZATION FLUOROIMMUNOASSAY; AMPHETAMINE; IMMUNOASSAY; URINE; methamphetamine; ELISA; antibody purification method; coating tracer; optimization
ISSN
0253-6269
URI
https://pubs.kist.re.kr/handle/201004/143965
DOI
10.1007/BF02974041
Appears in Collections: