Heat transfer analysis of variable stiffness endoscope utilizing thermoplastic and thermoelectric device inside large intestine

Other Titles
열가소성수지와 열전소자를 이용한 가변강성 내시경의 인체 내부 열전달 분석
Authors
KIM SEUNGWONKim MyoungjuKIM JUYEONGPark Jaehyeon
Issue Date
2021-05-20
Publisher
한국로봇학회
Citation
제16회 한국로봇종합학술대회
Abstract
Variable stiffness mechanism is required for the large intestine endoscopy to prevent medical accident and to reduce procedure time. The variable stiffness endoscope can provide two state that one is a stiff state for physical manipulation of tumor and the other is a soften state for navigation inside a serpentine path of large intestine. This paper analyzes the heat transfer effect of a variable stiffness mechanism utilizing a thermoplastic and thermoelectric device inside the large intestine by finite element method simulation. The result shows that the heat generated by thermoelectric devices of variable stiffness mechanism is higher than body temperature, the temperature at a contact point between the mechanism and large intestine does not rise rapidly due to the heat dissipation through the body.
Keywords
가변강성; 내시경; 열전달 분석